The Ebro FOM330 is widely used across the hospitality and food service sectors, the demand for accurate, rapid results and ease of operation is very capably fulfilled.
The monitor provides an accurate control of oil usage influencing continuity in texture, taste, and appearance of finished product whilst avoiding significant health issues arising from over-use of oil.
Polar compounds, now accepted as a reliable measurement of ageing oils, are assessed by the FOM330 oil monitor and displayed in digital results from zero to fifty.
Along with all the design development carried forward from previous models, this instrument features:
- Auto 'Hold'-release function - the instrument automatically releases the Hold function when it senses that it has been returned to hot oil. This eliminates the possibility of the operator transferring frozen readings from vat to vat.
- Increased digit size in display for greater visibility.
- Large oil quality signal lamp on top of the instrument gives a bright unmistakable indication of when oil is to be changed.
- 'Next Step' indicator in display - presenting clear yet unmistakably instructions to the user of what to do next.
- Improved temperature compensation for even greater accuracy in all situations.
- Larger ergonomic protective cover for increased protection from hot oil splashes.
- Improved carrying case.
Annual calibration facilities are available with a fully automated recall reminder service. Fast turnaround in our in-house oil testing laboratory ensures minimum downtime. Click here for more information.
Accuracy | Typically ±2.0 polar compound, ±1.0°C temperature | Dimensions | 125 x 54 x 22mm |
---|---|---|---|
Resolution | 0.5 polar compound, 1.0°C temperature | Sensor | Di-electric constant polar compounds. Temperature - thermistor |
Battery life | 5 years | Warranty | 2 years |
Range | 0…40% polar compound, 50…200°C | Protection | IP67 |
Instruction Manual
How often does my FOM330 need to be calibrated?
Your FOM330 needs to be calibrated annually to ensure reliable readings are maintained. To arrange a calibration, visit our calibration portal or email us at service@klipspringer.com.
How long will my FOM330 typically last for?
Your FOM330 has a two year guarantee, but if cared for and used in the correct way, FOM330s typically have a lifespan of around 5 years. You can find tips on how to care for your FOM330 here.
Does my FOM330 record temperature?
Yes your FOM330 does record temperature. This is a useful way to tell if your cooker is working in line with its temperature measurements.
Is there a direct link between Total Polar Compound (TPC) and acrylamide levels?
Whilst it is recognised the older your oil is, the higher the levels of TPC and acrylamide levels are, there has been no scientific evidence to show a direct correlation between the two levels. Find out more about reducing the risks of acrylamide here.
Is it still beneficial to single outlets?
It is beneficial for any individual kitchen. Some chains choose to adopt it to give them consistency from site to site as well as significant cost savings, however; many single kitchens adopt it as a great way to reduce oil spend.
What do the error messages (ER01/02/03) on my FOM330 mean and how should I respond?
Error Message ER01: Defective temperature sensor = send unit for service/repair
Error Message ER02: Temporary short circuit of the oil sensor = clean the sensor
Error Message ER03: Defective oil sensor = send unit for service/repair