Got a question? Speak to our industry experts on 01473 461800

PCOND210-K Conductivity Meter Kit with Electrode

Measures conductivity, TDS, resistivity, and salinity. Supplied with a plastic-body electrode, integrated temperature sensor, and a titanium-platinum black conductivity cell (k=1.0). Includes a carrying case and four conductivity solutions: 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, and 111.8 mS/cm (60 ml each). 

Calibration Certificate included. 

IP67 waterproof, shock- and scratch-resistant housing, combined with digital sensors for a flexible and dynamic measurement experience.

Code: PCOND210-K
770.45

1 Available


Code: PCOND210-K
Availability 1
770.45

Add to Favourites

Key Features:

  • Automatic (up to 4 points) and manual (up to 5 points) conductivity calibration.
  • Adjustable reference temperature and temperature coefficient.
  • Adjustable cell constant and selectable measurement unit.
  • Automatic / manual temperature compensation and temperature calibration function.
  • Conductivity and total dissolved solids (TDS) conversion.
  • Four TDS calibration curves for wide range of applications: EN27888, Linear, 442, NaCl.
  • Two salinity calibration curves: NaCl, Seawater.
  • °C & °F temperature readings.
  • Auto log data: built-in memory can store up to 500 data.
  • Auto stable, auto hold, and real-time measurement modes.
  • Auto shut-off (programmable: up to 30 minutes).
  • Monochrome LCD with white LED backlight: 50 x 50 mm.
  • Powered by 2 x AA batteries.
  • Casing is IP67 waterproof/dustproof, shock & scratch resistant, non-slip.
  • Packaged with electrode and calibration solutions in carrying case.
  • 3 years meter warranty.


Measurement Specifications

Conductivity


Range: 0 μS/cm to 200.0 mS/cm (k=1.0)

Resolution: 0.05% full scale

Accuracy: ±0.6% full scale, ±1.5% full scale > 18.0 mS/cm

Reference Temperature: 15 to 30 °C

Temperature Coefficient: 0.00 to 10.00%

Cell Constants: 0.1, 1.0, 10.0

Calibration Points: Up to 4 (Auto) / Up to 5 (Manual)

Units: S/cm, S/m (Auto Ranging)


Total Dissolved Solids (TDS)


Range: 0 ppm to 100 ppt (TDS factor=0.5)

Resolution: 0.01 ppm (mg/L) / 0.1 ppt (g/L)

Accuracy: ±0.1% full scale

TDS Curves: Linear (0.40 to 1.00), EN27888, 442, NaCl


Resistivity


Range: 0.000 Ω•cm to 20.0 MΩ•cm

Resolution: 0.05% full scale

Accuracy: ±0.6% full scale, ±1.5% full scale > 1.80 MΩ•cm


Salinity


Range: 0.0 to 100.0 ppt / 0.00 to 10.00 %

Resolution: 0.1 ppt / 0.01%

Accuracy: ±0.2% full scale

Salinity Curves: NaCl, Seawater

Calibration Option: Yes


Temperature


Range: -30.0 to 130.0 °C / -22.0 to 266.0 °F

Resolution: 0.1 °C / °F

Accuracy: ± 0.5 °C / ± 0.9 °F

Calibration Option: Yes


Package Details


PCOND210 Cond/TDS/Res/Sal/Temp meter

2 x AA batteries

Manual and quick guide

PCONDSEN210 plastic-body, Ti/Pt black conductivity probe (k=1.0) with integrated temperature sensor

84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm conductivity standard solutions (60ml each)

Carrying case

Memory500Auto Data Log YES
Auto Hold / Auto Stable/ Real Time YESAuto Shut-Off (1 to 30 mins.) YES
Electrode Status YESDiagnostic Messages YES
Software Upgrade1 YESMeter Inputs BNC, phono
DisplayCustom LCD with backlight HousingIP67, shock & scratch resistant, non-slip
Power Requirement 2 x AA batteries Battery life> 500 hours
Meter Dimensions 160 (L) x 80 (W) x 40.60 (H) mm Meter Weight Approx. 260 g (with batteries) / 216 g (without batteries)