The only pocket meter that directly measures conductivity in 0.12ml sample. No need for a beaker to calibrate your meter or measure a sample. Just place few drops of the standard or sample onto the sensor. This procedure saves you time and prevents wasting your precious sample.
The only pocket meter that directly measures conductivity in 0.12ml sample. No need for a beaker to calibrate your meter or measure a sample. Just place few drops of the standard or sample onto the sensor. This procedure saves you time and prevents wasting your precious sample.
Features:
- Minimum sample volume: 0.12ml
- Up to 3 calibration points
- Conductivity and temperature display
- Auto range
- Auto calibration
- Auto standard recognition: 1413µS/cm, 12.88mS/cm, 111.8mS/cm
- Auto temperature compensation: 2%/ºC fixed
- Temperature calibration: 5 to 40.0 ºC
- Auto stable / auto hold measurement
- Auto power off: 30 mins.
- Low battery indicator
- IP67 Rating dust & waterproof
- Replaceable sensor
- 400 Hours battery life: continuous use without backlight
- Materials: ABS epoxy body & titanium coated with platinum black sensor
- Display: custom (monochrome) LCD with backlight
- Operating conditions: 5 to 40°C, 85% or less relative humidity
- Warranty: 2 years - meter, 6 months - sensor
- Packed in handy carrying case
- Supplied with 1413µS/cm & 12.88mS/cm conductivity standards (14ml each), conditioning solution (4ml), 2 x CR2032 batteries, dropper, instruction and quick manuals
Accuracy | ±2% full scale (0 to 19.99 mS/cm), ±5% full scale (20.0 to 199.9 mS/cm) | Battery life | Approx. 400 hours in continuous use (without backlight) |
---|---|---|---|
Conductivity calibration points | Up to 3 | Conductivity Range & Resolution | 0 to 199 µS/cm: 1µS/cm, 200 to 1999 µS/cm: 1µS/cm, 2.00 to 19.99 mS/cm: 0.01 mS/cm, 20.00 to 199.9 mS/cm: 0.1 mS/cm |
Dimensions | 164 × 29 × 20mm (excluding projections) | Display | Custom (monochrome) LCD with backlight |
Mass | Approx. 50 g (including sensor and batteries) | Measurement Principle | 2 Electrode Bipolar AC |
Minimum Sample Volume | 0.12ml | Operating Humidity | 85% or less relative humidity (no condensation) |
Operating temperature | 5 to 40°C | Power | CR2032 batteries (2) |
Temperature Calibration | 5 to 40.0 ºC | Temperature Display | 0 to 50.0 °C |