The new Titrator TitroLine 5000 offers even more features than its predecessors and is even more convenient to use.
- High resolution pH/mV-measurement input for pH-, ORP-, silver- and further mV-electrodes
- Pt1000 and NTC 30 temperature measurement input for automatic temperature compensation
- Linear and dynamic titration to equivalence points
- Determination of acidity/acid degree in milk and milk products.
- Determination of total acidity in fruit juice.
- Salt contents in food samples.
- pH- value, total acidity in wine and beverages.
- Chloride in drinking water and other samples
Measurement input pH/mV | pH/mV-input with Electrode socket according to DIN 1962 or additional with BNC socket insert (Z 860) | Measurement input temperature | Pt 1000/NTC 30: (socket 2 x 4 mm) |
---|---|---|---|
Interfaces | 1 x USB-A and 1 x USB-B, 2 x RS-232-C | Stirrer connection | TM 50 power supply directly through piston burette |
Keyboard | The unit is operated using the keys on the device itself, the controller TZ 3880 and optional PC-keyboard (USB) | Display | graphics-capable TFT display |
Volume display | 0000,000...9999,999 ml | Display resolution | 0.005-0.025 ml (depending on dosing unit) |
Dosing speed | max. 100 ml/min (with 50 ml unit) | Filling Speed | min 30 s to 999 s adjustable (time according to the cylinder volume) |
Dosing units | 20 ml or 50 ml dosing unit, interchangeable | Burette resolution | 8000 |
Dosing accuracy | systematic error 0.15 %, random error 0.05% in compliance with EN ISO 86556 | Power supply | 100 -240 V~; 50/60 Hz, power input 30 VA |
Conformity | ISO 8655, part 6 | CE-mark | EMC: 2004/108/EG; safety EG-Directive 2006/95 |
Dimensions | 135 x 310 x 205 mm (W x H x D), including dosing unit, without stirrer | Weight | 2.3 kg (without stirrer) |